0 40 80 120 160 180 200 UP/BP/ULP/SLP 400 - 800 OPERATING INSTRUCTIONS UP 400 – 800 BP 400 - 800 ULP/SLP 400 – 800 OPERAT
10 UP/BP/ULP/SLP 400 - 800 4 Main switch module The main switch module contains: main switch to select the working mode three signal lamp
UP/BP/ULP/SLP 400 – 800 11 5 Working controller module The working controller can be operated using the digital rotary knob and the set butt
12 UP/BP/ULP/SLP 400 - 800 6 The Memory Card XL 6.1 Programming: A temperature profile with up to 40 ramps can be programmed on the Memory Ca
UP/BP/ULP/SLP 400 – 800 13 6.3 Programming with the PC program "CELSIUS 2000" Connect the PC to the oven via a shielded serial inter
14 UP/BP/ULP/SLP 400 - 800 7 The parallel printer interface All units of the class P are fitted with a parallel interface of the same type as u
UP/BP/ULP/SLP 400 – 800 15 Printer mode The menu item PRINT activates the data transmission to the printer. Numerical print numerical printo
16 UP/BP/ULP/SLP 400 - 800 8 Serial communication interface 8.1 Communication protocol The communication protocol of the MEMMERT P-controller
UP/BP/ULP/SLP 400 – 800 17 8.4 Interface RS485 If specially ordered, the unit can be fitted at the works with an RS485 interface instead of an
18 UP/BP/ULP/SLP 400 - 800 9 Working controller module with main switch in position I (continuous operation) With main switch in position I the
UP/BP/ULP/SLP 400 – 800 19 Menu item TEMP to adjust the set temperature. FAN serves to
2 UP/BP/ULP/SLP 400 - 800 TABLE OF CONTENTS 1 Congratulations on your choice of a MEMMERT Oven!...4 2
20 UP/BP/ULP/SLP 400 - 800 9.1 Submenu SETUP After selecting YES under the menu item SETUP you branch to the submenu for the configuration of th
UP/BP/ULP/SLP 400 – 800 21 time in 24-hour format setting range: 00.00h to 23.59h date in the format day-month. sett
22 UP/BP/ULP/SLP 400 - 800 9.2 The working controller module with main switch in position After switching on, the display shows for a few sec
UP/BP/ULP/SLP 400 – 800 23 Note ! During programme run (indicated by the flashing of the LED prog) it is impossible to change the selected par
24 UP/BP/ULP/SLP 400 - 800 9.3 Editing the programme / EDIT submenu The programming submenu is opened by selecting YES in the menu item EDIT (
UP/BP/ULP/SLP 400 – 800 25 The dynamic operating menu only shows as many programme segments as programmed, i.e. until a programme segment is clo
26 UP/BP/ULP/SLP 400 - 800 9.4 Segment end command The programme segments are connected by a segment end command. The end commands determine th
UP/BP/ULP/SLP 400 – 800 27 10 Programme start A programmed temperature profile can be started as described in the chapter Controller Module wit
28 UP/BP/ULP/SLP 400 - 800 10.1 Programming example An oven with fan has to be heated to 100°C in one hour, with the air flap closed and with
UP/BP/ULP/SLP 400 – 800 29 ϑ100°C20°C1htFANFLAP2h 15min1h
UP/BP/ULP/SLP 400 – 800 3 14 Sterilisers ...
30 UP/BP/ULP/SLP 400 - 800 10.2 Programming example for setpoint-dependent operation An oven with fan has to heat up to 180°C in one hour, wi
UP/BP/ULP/SLP 400 – 800 31 ϑ180°C20°C1htFANFLAPset temperatureactual temperaturewait2h 15min1hhold
32 UP/BP/ULP/SLP 400 - 800 11 The MEMMERT software „Celsius 2000“ Using the „Celsius 2000“ program: the P-controller can be programmed from
UP/BP/ULP/SLP 400 – 800 33 12 Safety devices 12.1 Adjustable overheat controller (TWW), Class 3.1 in accordance with DIN 12880 or adjustable te
34 UP/BP/ULP/SLP 400 - 800 Should the temperature rise above the set working temperature, The adjustable overheat controller TWW takes over th
UP/BP/ULP/SLP 400 – 800 35 13 Loading Units described in this instruction manual must not be used for drying or heating substances giving off
36 UP/BP/ULP/SLP 400 - 800 14 Sterilisers 14.1 Application The apparatus is used for sterilising medical materials by employing dry heat through
UP/BP/ULP/SLP 400 – 800 37 In particular with heavy loads, these approximate values should not be used without proper testing. In order to ensur
38 UP/BP/ULP/SLP 400 - 800 14.4 Example of programming a sterilisation profile The following example shows a possible sterilisation profile.
UP/BP/ULP/SLP 400 – 800 39 ϑ180°C20°C1mintFANFLAPset temperatureactual temperaturewait1 h1minwait1minEnd
4 UP/BP/ULP/SLP 400 - 800 The text of these Operating Instructions has been translated from the German. If any part of the text is doubtful or th
40 UP/BP/ULP/SLP 400 - 800 15 Maintenance General information Memmert ovens require minimal servicing. It is recommended that the moving parts o
UP/BP/ULP/SLP 400 – 800 41 16 Cleaning Regular cleaning of the (easy to clean) interior prevents the formation of deposits which may impair
42 UP/BP/ULP/SLP 400 - 800 The details in these Operation Instructions must be carefully observed in order to ensure satisfactory operation. Any
UP/BP/ULP/SLP 400 – 800 43 EC Declaration of Conformity Manufacturer´s name and address: MEMMERT GmbH + Co. KG Äußere Rittersbacher Straße
44 UP/BP/ULP/SLP 400 - 800 EC Declaration of Conformity Manufacturer´s name and address: MEMMERT GmbH + Co. KG Äußere Rittersbacher Straße 3
UP/BP/ULP/SLP 400 – 800 45 EC Declaration of Conformity Manufacturer´s name and address: MEMMERT GmbH + Co. KG Äußere Rittersbacher Straße
46 UP/BP/ULP/SLP 400 - 800 19 Index ...page ...
UP/BP/ULP/SLP 400 - 800 47 R Readjusting the door ...
D07981 01.10.2002
UP/BP/ULP/SLP 400 – 800 5 2 The equipment of MEMMERT cabinets Series UP, BP, ULP, SLP consists of: an electronic fuzzy-supported microproces
6 UP/BP/ULP/SLP 400 - 800 Technical data summary A Pt 100 DIN A in 4-wire circuit measures the temperature in the working chamber. The temperatur
UP/BP/ULP/SLP 400 – 800 7 model volume current power voltage ±10% weight UP/ULP/SLP 400 53 l 6,1A 1400 W 230 V~ 35 kg UP/ULP/SLP 5
8 UP/BP/ULP/SLP 400 - 800 2.2 Installation options (special equipment) The cabinet can be setup on the floor or on a table (bench), model
UP/BP/ULP/SLP 400 – 800 9 2.5 Stacking device (special equipment) Where two ovens of the same size are to be placed one on top of the other, t