WaterbathsWNB 7 - 45Operating instructiOns
BASIC page 104.3.1 Temperature setpointThe bath starts to heat up immediately to the set temperature.Adjustment range:10°C to 95°C or ca. 100°C withac
BASICpage 114.3.3 Hold time of the setpoint temperatureThe bath switches off the heating after the end of the selected hold time. In this case the ho
BASIC page 126 Monitoring the programme sequenceBath does not heat upDuring the delayed switch-on the symbol flashes and the display alternates betw
BASICpage 137 Temperature monitor and protection devices7.1 Mechanical temperature monitor: temperature limiterAll waterbaths are fitted with a mec
BASIC page 148 Use of the cover (special equipment)Gable cover The gable cover (may be ordered as already mounted special equipment) should always b
BASICpage 1510 Cooling system (special equipment)If the waterbath is fitted with a cooling device for quicker cooling of the bath liquid, the “water
BASIC page 1612 Check list for fault rectificationIn case of a malfunction contact an authorized service station for Memmert equipment or please in
BASICpage 1715 CE Conformity of DeclarationEC Declaration of Conformity Manufacturer´s name and address: MEMMERT GmbH + Co. KG Äußere Rittersbache
BASIC page 18Standard units are safety-approved and bear the test marks:This product is subject to the Directive 2002/96/EC by the European Parliament
BASICpage 1917 IndexAaddress 18alarm symbol 13ambient conditions 4ambient temperature 16autodiagnostic system 5, 16Bbath construction 8boiling m
BASIC page 21 General notes and safety notes ...
08.06.2009Waterbath BASIC englischD10329Memmert gmbH + co. Kg | Po. Box 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 14
BASICpage 31 General notes and safety notesYou have purchased a technically fully proven product which has been produced in Germany with the use of h
BASIC page 42 Technical dataModel 7 1014222945Volume [liter] 7 10 14 22 29 45Usable bath length A [mm] 240 350 350 350 590 590Usable bath width B [mm
BASICpage 52.1 Material qualityMemmert is using stainless steel ( Mat.Ref. 1.4301) for the external casing as well as for the interior, an outstandin
BASIC page 62.5 Standard equipment• Electronic fuzzy-supported PID process controller with delayed programme start and programmable hold time. Th
BASICpage 73.1 Initial start-upWhen the bath is started up for the first time, it should be supervised continuously until steady conditions have been
BASIC page 84 Bath construction and operationThe heating positioned on three sides around the tank ensures a natural water circulation of the liqui
BASICpage 94.2 Switching onThe bath is switched on by pressing the push/turn controland can be operated in connection with the SET-key.Bath switched